Product Introduction
SY/T 7329-2016 Oilfield Chemical AgentOrganic Chlorine Content Analyzer is designed and manufactured based on the oxygen combustion method for determining organic chlorine content in oilfield chemical agents. The sample is decomposed by oxygen combustion to convert organic chlorine into inorganic chlorine. After absorption by sodium hydroxide solution, the total chloride ion content is measured using this instrument. The inorganic chloride ion content in the oilfield chemical agent is then determined. The difference between the total chlorine content and the inorganic chlorine content represents the organic chlorine content in the oilfield chemical agent. It is suitable for the determination of organic and inorganic chlorine in oilfield chemical agents and can be widely used in power, petroleum, chemical, commercial inspection, and scientific research sectors.
Features
◆Complies with: SY/T 7329-2016 "Determination of Organic Chlorine Content in Oilfield Chemical Agents" Method B
◆Complies with: Q/SH1020 2093-2016 "Determination of Organic Chlorine Content in Oilfield Chemical Agents" Method A
◆Windows operating platform, user-friendly interface with workstation functionality, simple operation, reliable performance, excellent stability, and easy installation
◆Suitable for all methods using silver nitrate titration for chloride ion determination. Enables accurate measurement with higher sensitivity and simpler operation without the need for solution calibration. Suitable for both high-content chloride ion determination and trace chloride ion detection in samples
◆Computer-controlled, uses a titration cell with a specially designed sintered glass disc replacing the ion membrane, avoiding erroneous results due to membrane damage
◆Microcomputer-controlled, displays the entire process status and parameters, prints various results. Multi-parameter setting and adjustment make analysis of different sample types simpler, faster, with shorter analysis time and more accurate results
◆Complete instrument includes: main unit, electrolytic cell, stirrer, oxygen combustion bottle, computer, etc.
◆Supports full Chinese/English interface display (optional)
Technical Parameters
Applicable Standards | SY/T 7329-2016, Q/SH1020 2093-2016 |
Sample Types | Oil-based, water-based, solid |
Measurement Method | Coulometric method |
Bias Voltage Range | 0~500mv |
Measurement Range | Chloride ion content: 0.1ppm to percentage content |
Sensitivity | Cl-: 0.1mg/L |
Accuracy | <±2% |
Repeatability | ≤2mgNaCl/L ±0.1mgNaCl/L >2mgNaCl/L ±5% |
Sample Amount | 0.03 or 0.05g |
Power | 50W |
Ambient Temperature | 5℃~45℃ |
Relative Humidity | 10%~80%Rh |
Power Supply | AC220V±10% 50Hz |
| Industry Category | Measurement-Analysis-Instruments |
|---|---|
| Product Category | |
| Brand: | 旭鑫仪器 |
| Spec: | ST-1530D |
| Stock: | |
| Origin: | China / Beijing / Changpingqu |